Aerosol Photometers have continued to grow in application and utility, particularly in the last decade. With their increasing application to ambient environmental monitoring and Homeland Defense applications, minimal maintenance has become of paramount importance. The obvious principle of blowing air over the lenses does not work. Sufficient air to keep the lens truly clean amounts to an air flow rate equal to the main flow through the instrument. This in turn leads to erroneous readings which will vary with concentration and aerosol properties.


By structuring and containing the flow of the sample through the light beam and ensuring that it is within the laminar flow regime we are able to cause the stream of aerosol laden gas to “jump” the light beam gap. Hence there is no dispersion of the aerosol particles and no lens deposition. This arrangement is simply illustrated in the accompanying drawing:

This BGI Patented device has been designed to fit the Casella Microdust Pro Photometer. When used with any non pulsating flow rate, less than 16 lpm all aerosol particles will “jump the gap” and lenses will remain clean.
(covered under patent number 7,256,882 B2)

 

Applications

  • Active air flow through the Microdust Pro
  • Long period concentration measurements
  • Toxicology monitoring applications
  • Homeland Defense monitoring
  • Aerosol research

Features

  • Adapts to all models of Microdust Pro
  • Adapts to previous AMS models with same probe diameter
  • Uses most personal and area samplers for air flow
  • Simple to install/detach
  • Prolongs lens cleanliness indefinitely

Downloads

Download the latest version of the instruction manual. Click here and then click on manual # 32.

Specifications

Inlet outside diameter 1/2" (1.27 cm)
Inlet inside diameter 3/8" (0.96 cm)
Inlet length 11/16" (1.75 cm)
Outlet Standard or Slaton 37mm Cassettes
Overall Dimensions 1.62" (4.11 cm) x 2.00" (5.08 cm) x 6.5" (16.5 cm)
Weight 4.74 oz. (296 grams)
Mounting points (2) #1/4-20 (1/4 Whit) holes provided on narrow face 2.0" apart

 

Ordering Information
Click on a part number to add that item to your Price Quote List.
LFD2 Clean lens adaptor
 
Replacement parts
LFD126BUNA Main tube "O" rings 6/pkg
LFD015SIL Inlet/outlet "O" rings 6/pkg
BGI4.4 Cassette "O" ring 6/pkg