deltaCal Calibrator for FRM Samplers

   

 

deltaCal Control Module and Measuring Head
deltaCal Control Module and Measuring Head

 

 

 

(click on image below for a larger detailed image)
deltaCal Display Screens
deltaCal Display Screens

 

 

 

 

deltaCal Complete in Carrying Case
deltaCal Complete in Carrying Case

 

 

 

 

 

Sectional View of Measuring Head
Sectional View of Measuring Head
(Click on Image to Enlarge)

 

 

 

 

 

Typical Computer Screen - deltaCal Software
Typical Computer Screen - deltaCal Software
(click on image to enlarge)

Designed to go anyplace, indoors or out. From the top a 15,000 ft mountain to a 55° C desert or a -30° C snow field. Our first principle, Venturi calibrators will operate for 20 continuous hours on 4 AA cells. We are pulsation free, accurate to less than 0.75% of the flow rate reading, NIST traceable and designed, built and calibrated in an ISO 9001-2000 facility. We have always built our calibrators to these standards. Now our output screens show both Actual and Standard flow rates. This feature is applicable to every calibrator that we have ever produced and will be applied, during a recalibration, to every instrument at no charge.

The BGI deltaCal is the first and only continuous volumetric flow calibration/audit device which gives a direct indication of volumetric flow rate, standard flow rate, barometric pressure (both mm/Hg & mb), ambient temperature and filter temperature in one instrument. It was developed for use with U.S. EPA PM2.5 designated samplers and their PM10 derivatives. With an accessory adapter it is usable with any air sampler in the 1.5 to 19.5 Lpm range. The DC100 version has a range of 10-60 Lpm. Typical screens are shown below on the left.

It is the only field audit/calibration device which meets the same temperature criteria as the EPA designated samplers themselves. The temperature range is -30° to 55° C (EPA requires up to 40° C).

The heart of the deltaCal is a venturi which is superior to an orifice, in that it offers no sharp edges which can be damaged and a 90% recovery of the measurement pressure drop, important when calibrating some forms of speciation samplers or other critical orifice controlled devices. Because flow through the instrument is uninterupted, our graphic software provides a continuous output display of all functions.

Applications

  • Calibration and audits of PM2.5 and PM10 samplers to U.S. EPA requirements
  • May be used on any sampler in the 1.5 to 19.5 Lpm range
  • Usable in field or laboratory
  • Calibration of European PM10 samplers operating at 38.4 Lpm
  • Calibration of 50 Lpm DICOT samplers
  • Calibration of standard DICOT samplers.

Features

  • One switch (off/on) operation -- all data displayed on screen
  • No moving parts, completely silent.
  • 20 hour battery life (4-AA)
  • Furnished with universal power supply (120/240 VAC)
  • Download feature with supplied software

Downloads

deltaCal Specifications

Flow Rate Range 1.5-19.5 Lpm (Euro 10-60 Lpm)
Temp. operational range -30° C to 55° C
Temp. reading range -40° C to 55° C
Baro. pressure range 400 to 800 mm of Hg (U.S. version)
534 to 1066 millibars (Euro version)

Dimensions

Measuring Head 5 in. Max OD (12.7 cm) X 10 in. high (25.4 cm)
Control Module 8 in. high (20.3 cm) X 4 in. wide (10.15 cm) X 1.7 in. thick (4.32 cm)
Combined Weight 2.2 lbs (1 kg)

Hand-Held Temperature Probe

Length of Cable 33 in (83.8 cm)
Length of Probe 5 in. (12.7 cm)

Carrying Case

Dimensions 18 in. wide (45.7 cm) X 13 in. high (33 cm) X 7 in. thick (17.8 cm)
Weight Complete with Contents 5.5 lbs (2.5 kg)

Ordering Information
Click on a part number to add that item to your Price Quote List.

DC1 deltaCal (calibration range 1.5-19.5 Lpm), Complete in Carrying Case, Std. Cond. Temp at 25° C
DC1MB deltaCal (calibration range 1.5-19.5 Lpm), Complete in Carrying Case, Std. Cond. Temp at 0 ° C
DC100 deltaCal (calibration range 10-60 lpm), Complete in Carrying Case , Std. Cond. Temp at 25° C
DC100MB deltaCal (calibration range 10-60 lpm), Complete in Carrying Case, Std. Cond. Temp at 0 ° C

Accessories

DC10 Adapter - deltaCal to 3/8 in. hose barb
(use with SASS or TEOM)

Replacement Parts

DC2 Extension Cable Temperature Probe
DC3 120/240V Power Supply
DC4 Fitted Carrying Case
DC6 Pressure tap "O" rings (4)
DC7 Venturi Thumb screw - ea
X010 Replacement socket "O" rings (2)

Recalibration

DC12 Recalibration with NIST Traceability Certificate